recall_number
string
recall_type
string
firm_name
string
product_description
string
reason_for_recall
string
status
string
recall_class
null
initiation_date
date32
report_date
date32
distribution_pattern
string
product_quantity
string
product_ndc
string
fei_number
string
facility_total_inspections
float64
facility_oai_count
float64
facility_last_classification
string
facility_last_inspection
date32
facility_compliance_actions
int64
facility_warning_letters
int64
last_compliance_action_date
date32
preceded_by_oai
bool
days_since_last_inspection
float64
failure_category
string
failure_subcategory
string
D-0273-2025
drug
Denison Pharmaceuticals, LLC
Zapzyt, Acne Treatment Gel, 10% benzoyl peroxide Gel, packaged in 1 oz (28.35g) tube, Distributed by: FOCUS CONSUMER HEALTHCARE, LLC., 801 Broad St Ste 200, Chattanooga, TN 37402. NDC 71687-0011-1
Chemical Contamination: Presence of benzene.
Ongoing
null
2025-02-24
2025-03-19
Nationwide within the USA
642,131 tubes
71687-0011
3012228456
1
0
Voluntary Action Indicated (VAI)
2016-04-05
0
0
null
false
0
contamination
benzene presence
Z-0472-2017
device
ICU Medical, Inc.
RIGHT HEART KIT FOR NC BAPTIST HOSPITAL, Item No. 46066-34 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure ...
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
Terminated
null
2016-09-27
2016-11-23
null
125 units
null
3013319212
8
2
No Action Indicated (NAI)
2024-09-09
2
2
2013-05-09
false
-1
device_malfunction
potential leak
Z-1187-2026
device
Medline Industries, LP
Medline medical procedure convenience kits labeled as: 1) CATH LAB PACEMAKER PACK-LF, Kit SKU DYNJ64837C; 2) LOOP RECOVERY PACK, Kit SKU DYNJT7034;
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Ongoing
null
2025-12-12
2026-02-04
null
1,928 total
null
3015077251
1
0
No Action Indicated (NAI)
2024-04-24
0
0
null
false
0
device_malfunction
tubing cracking defect
Z-2553-2014
device
Siemens Healthcare Diagnostics, Inc.
Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Terminated
null
2014-07-16
2014-09-03
null
2757
null
1219913
8
1
Voluntary Action Indicated (VAI)
2022-03-08
1
1
2012-05-10
false
-1
device_malfunction
hinge failure risk
D-1595-2019
drug
Herbal Doctor Remedies
Dragon Pain Oil, Distributed by Herbal Doctor Remedies Alhambra, CA 91801-7081 USA, 1(800) 600-0808
CGMP Deviations: Products were manufactured outside of the controls required by current Good Manufacturing Practices.
Terminated
null
2019-07-12
2019-08-14
Nationwide within the United States and Brazil and on the internet
N/A
null
3003040962
3
1
No Action Indicated (NAI)
2018-05-22
2
2
2019-10-04
false
0
cGMP
manufacturing controls
Z-0836-2018
device
Stryker Corporation
STRYKER CASSETTE PUMP, AHTO Tube Set, labeled sterile. Includes the following: a. Stryker AHTO Tube Set Packaging (Model 0250070600); b. Stryker AHTO Tube Set with Tip Packaging (Model 0250070620);
Routine bioburden testing of certain lots were found to have levels higher than internal acceptable rates.
Terminated
null
2017-11-17
2018-03-14
null
60,753 units total
null
3006011135
1
0
Voluntary Action Indicated (VAI)
2010-09-24
0
0
null
false
0
sterility
elevated bioburden levels
Z-0493-2015
device
Bard Access Systems
GuardIVa Antimicrobial Hemostatic IV Dressing, REF FP-23-AD008, Manufacturer: Bard Access Systems, Inc. USA. Product Code FP23AD006.
Bard Access Systems is recalling GuardIVa (Ref. No. FP23AD006) from lot ASXHT003 because they were non-sterile unit samples which potentially could be used clinically.
Terminated
null
2014-11-03
2014-12-10
null
2532 units
null
null
null
null
null
null
0
0
null
false
null
sterility
non-sterile units
Z-0001-2019
device
Smith & Nephew, Inc.
Journey BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 74021222, 74021223, 74021224, 74021225, 74021226, 74021227, 74021228, 74021229
The data in the National Joint Registry of England, Wales and Northern Ireland (NJREWNI) and Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) indicate that the first generation JOURNEY BCS Knee System has a revision rate over 1.5 times the primary total knee arthroplasty device class ave...
Ongoing
null
2018-06-13
2018-10-10
null
10,792 units
null
3007282753
2
0
No Action Indicated (NAI)
2017-07-24
0
0
null
false
0
device_malfunction
elevated revision rate
Z-3397-2018
device
B. Braun Medical, Inc.
EC19CF 19G X 100CM CATH CLOSED TIP LF, Material Number 333512 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous admini...
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Terminated
null
2018-07-20
2018-10-10
null
16400
null
3014301563
1
0
Voluntary Action Indicated (VAI)
2018-12-27
0
0
null
false
-1
device_malfunction
connector closure failure
Z-1328-2017
device
Windstone Medical Packaging, Inc.
Lap Pack, part number AMS4801
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
Terminated
null
2017-01-11
2017-03-08
null
36 kits
null
1000125955
2
0
No Action Indicated (NAI)
2019-06-21
0
0
null
false
-1
sterility
compromised glove integrity
Z-0211-2026
device
Cardinal Health 200, LLC
Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodena...
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Ongoing
null
2025-09-18
2025-10-29
null
all lots
null
3002973487
3
0
No Action Indicated (NAI)
2015-07-20
0
0
null
false
0
device_malfunction
valve breakage
Z-0680-2024
device
Abbott Laboratories
Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the...
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
Ongoing
null
2023-10-19
2024-01-17
null
1,325 systems
null
3013324964
9
0
No Action Indicated (NAI)
2025-08-01
0
0
null
false
-1
undeclared_allergen
latex allergen
Z-0471-2026
device
Medline Industries, LP
Medline Renewal LigaSure Blunt Tip Sealer/Divider, Compatible with FT10 Generator, 5 mm x 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.
Ongoing
null
2025-10-01
2025-11-19
null
2 units
null
3011508621
1
0
Voluntary Action Indicated (VAI)
2018-05-25
0
0
null
false
0
device_malfunction
material entrapment
Z-1881-2013
device
Siemens Healthcare Diagnostics, Inc
ADVIA Centaur Systems TSH3 Ultra ReadyPack ((REF 06491072 - 100 Tests, REF 06491080 - 500 Tests, REF 04862625 - 2500T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).
A rare variant of TSH, identified in a small cluster of patients, is not detected.
Terminated
null
2013-05-10
2013-08-14
null
172913
null
null
null
null
null
null
0
0
null
false
null
potency
undetected TSH variant
Z-1331-2023
device
Medtronic Sofamor Danek USA Inc
Medtronic Grafton DBF Inject: a) GRAFTON T50303 DBF INJECT 3CC, REF T50303; b) GRAFTON T50306 DBF INJECT 6CC, REF T50306; c) GRAFTON T50309 DBF INJECT 9CC, REF T50309; Demineralized Bone Matrix (DBM) product
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Ongoing
null
2023-02-08
2023-04-12
null
10037 units
null
1000477302
11
0
Voluntary Action Indicated (VAI)
2015-11-03
0
0
null
false
0
packaging
sterile barrier breach
Z-1344-2014
device
LumiQuick Diagnostics Inc.
Rotavirus Test Card LumiQuick, Santa Clara, CA 95054
Products shipped to US locations that do not have approved applications for PMA , IDE or 510(k) clearance.
Terminated
null
2014-03-04
2014-04-16
null
ALL
null
3005984081
5
1
No Action Indicated (NAI)
2025-11-05
2
2
2014-04-01
false
-1
other
unapproved device distribution
Z-2157-2012
device
Abbott Point Of Care Inc.
I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.
The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.
Terminated
null
2012-03-01
2012-08-15
null
2,950,296 cartridges US; 156,408 cartridges OUS
null
2245578
2
0
No Action Indicated (NAI)
2017-09-20
0
0
null
false
-1
device_malfunction
incorrect test results
Z-2553-2014
device
Siemens Healthcare Diagnostics, Inc.
Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Terminated
null
2014-07-16
2014-09-03
null
2757
null
3009448641
1
0
No Action Indicated (NAI)
2017-11-29
0
0
null
false
-1
device_malfunction
hinge failure risk
Z-2164-2012
device
Radiometer America Inc
ABL90 FLEX software version 2.7 MR3. The ABL90 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood.
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL...
Terminated
null
2012-07-05
2012-08-15
null
1427 ABL 90 Instruments
null
1523456
2
0
Voluntary Action Indicated (VAI)
2015-03-02
0
0
null
false
-1
software
HL7 data transmission
Z-0674-2013
device
Philips Medical Systems (Cleveland) Inc
Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Scanning Systems: Brilliance 64 - Model #728231; and Ingenuity CT - Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body...
Software anomalies. Philips Medical Systems recalled their Brilliance 64 running software v4.0.0xx516 and their Ingenuity CT running software v4.0.0xx518 due to software issues detected in these CT scanning systems. If these issues were to re-occur they could pose the risk of injury to patients undergoing scans; devi...
Terminated
null
2012-12-03
2013-01-23
null
A total of nine (9) units were distributed (3 units in the U.S & 6 units overseas)
null
1525965
10
2
No Action Indicated (NAI)
2019-11-26
1
1
2011-06-07
false
-1
software
CT scanner anomalies
Z-2427-2025
device
Boston Scientific Corporation
WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is u...
Ongoing
null
2025-07-29
2025-09-10
null
340185 units in total
null
3002095335
11
0
No Action Indicated (NAI)
2025-03-18
0
0
null
false
0
device_malfunction
air embolism risk
Z-3227-2024
device
American Contract Systems, Inc.
Endovascular AAA Pack, ESED52B; Medical convenience kit
ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Ongoing
null
2024-08-07
2024-10-02
null
93 units
null
3008493588
2
0
No Action Indicated (NAI)
2014-01-09
0
0
null
false
0
cGMP
equipment failure monitoring
Z-1669-2018
device
Siemens Healthcare Diagnostics, Inc.
Dimension¿ Free Thyroxine (FT4L) Flex¿ reagent cartridge
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Terminated
null
2018-01-02
2018-05-16
null
N/A
null
2517506
4
1
Voluntary Action Indicated (VAI)
2014-04-23
1
1
2012-06-29
false
0
mislabeling
incorrect instructions
D-0757-2022
drug
Pfizer Inc.
Accuretic (quinapril HCl/hydrochlorothiazide) Tablets, 20 mg/25 mg, 90 tablets bottles, Rx Only, Distributed by: Parke-Davis, Division of Pfizer Inc, NY, NY 10017, Made in Germany, NDC 0071-0223-23.
CGMP Deviations: N-nitroso-quinapril and N-nitroso-hydrochlorothiazide impurity above the acceptable daily intake limit.
Terminated
null
2022-03-21
2022-04-13
Nationwide
265 bottles
null
3001433984
2
0
No Action Indicated (NAI)
2025-08-22
0
0
null
false
-1
cGMP
N-nitroso impurity
Z-0528-2018
device
Angiodynamics, Inc.
Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602, Model # (UPN) H7879006020, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
Terminated
null
2017-06-05
2018-02-14
null
14,353 units in total
null
3017892510
10
0
No Action Indicated (NAI)
2021-06-25
0
0
null
false
-1
device_malfunction
coolant ingress failure
D-1011-2013
drug
Beacon Hill Medical Pharmacy, P.C.
Zinc Sulfate 1 mg/ml INJECTABLE, all configurations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
Terminated
null
2013-07-26
2013-09-18
USA nationwide
N/A
null
3009641885
1
1
Official Action Indicated (OAI)
2013-07-19
1
1
2014-09-24
true
0
sterility
QC procedure deficiency
Z-0701-2019
device
Boston Scientific Corporation
IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent ...
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Ongoing
null
2018-11-28
2019-01-16
null
1,282 (all three types in Japan)
null
1000116179
8
1
No Action Indicated (NAI)
2017-12-11
1
1
2014-04-10
false
0
cGMP
unapproved supplier change
D-335-2013
drug
Performance Plus Marketing, Inc.
BLUE Diamond PLATINUM Capsules, 1000 mg, supplied in 1, 5 and 10 count blister packs, Distributed By: DH Distribution Los Angeles, CA 90026 1 count blister UPC -608641933543; 5 count blister UPC - 608641932867; 10 count UPC -608641933536
Marketed Without an Approved NDA/ANDA: product may contain undeclared sildenafil, tadalafil and analogues of these FDA approved active ingredients, making them unapproved drugs
Terminated
null
2012-12-17
2013-05-22
Nationwide
1,000 capsules.
null
null
null
null
null
null
0
0
null
false
null
other
unapproved drug marketed
D-0095-2018
drug
NEW RELIANCE TRADING, INC.
No.1 Faiza Beauty Cream Manufactured by: Poonia Brothers (Pak), Gujranwala, Distributed by NEW RELIANCE TRADING, INC., 5563 59th St., Maspeth, New York, 11378-2358, UPC 8993138993349, 5842109854239.
Marketed Without an Approved NDA/ANDA
Terminated
null
2017-10-06
2017-12-06
NY
513 foil packets
null
null
null
null
null
null
0
0
null
false
null
other
unapproved marketing
Z-1318-2017
device
Windstone Medical Packaging, Inc.
Hystero Pack, part number PSS3535(A
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
Terminated
null
2017-01-11
2017-03-08
null
32 kits
null
1000125955
2
0
No Action Indicated (NAI)
2019-06-21
0
0
null
false
-1
packaging
compromised glove integrity
Z-3030-2017
device
CareFusion 303, Inc.
VersaSafe Plastic Cannula - 11 Gauge, Model 9391-0200 Product Usage: Used with secondary IV sets or syringes with a female luer lock adapter to administer fluid and medications through a needle or catheter inserted into a patients artery or vein.
Punctures in blister packaging that was detected during the packaging process.
Terminated
null
2017-06-30
2017-08-23
null
1,546,800 units
null
1034172
2
0
No Action Indicated (NAI)
2016-07-19
0
0
null
false
0
packaging
punctured blister pack
D-390-2014
drug
Physicians Total Care, Inc.
Carisoprodol IV ( Carisoprodol Tablets USP) 350mg, RX only. Mfd by: Shasun Chemicals Drugs LTD, Pondicherry, INDIA. Repacked and Distributed by: Physicians Total Care, Inc., Tulsa, OK 74146. 90 count bottle: NDC: 54868-0816-8, 60 count bottle: NDC: 54868-0816-4, 30 count bottle: NDC: 54686-0816-3.
Presence of Foreign Substance; heavy metals (chromium, titanium etc) and inactive components of the product were visually observed during routine stability testing.
Terminated
null
2013-11-22
2014-01-01
CA, GA, NC, and MO
3,000 tablets
null
1641527
3
1
No Action Indicated (NAI)
2014-03-18
2
2
2013-01-10
false
-1
foreign_material
heavy metals detected
D-0619-2025
drug
Amneal Pharmaceuticals, LLC
chlorproMAZINE Hydrochloride Tablets, USP, 200 mg, 100-count bottles Rx only, Manufactured by: Amneal Pharmaceuticals Pvt. Ltd., Oral Solid Dosage Unit, Ahmedabad 382213, INDIA; Distributed by: Amneal Pharmaceuticals LLC, Bridgewater, NJ 08807, NDC 69238-1062-1.
Presence of a foreign substance.A specific lot of auxiliary polyester coil, used in product packaging was detected with presence of a micro-organism. No micro-organism was detected on any tablets.
Ongoing
null
2025-08-18
2025-09-10
Distributed Nationwide in the USA
3,363 100-count bottles
69238-1054
3004488497
8
0
No Action Indicated (NAI)
2024-02-08
0
0
null
false
0
foreign_material
microorganism contamination
Z-0221-2017
device
Windstone Medical Packaging, Inc.
Laparotomy Kit, Kit number PSS1036(A and PSS1036(B convenience custom kits used for general surgery in hospital operating room
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES, Part # 13140, of the custom AMS kit were recalled because gloves may contain splits or holes compromising sterility.
Terminated
null
2015-04-22
2016-10-26
null
230 kits
null
1000125955
2
0
No Action Indicated (NAI)
2019-06-21
0
0
null
false
-1
device_malfunction
splits or holes
Z-3016-2020
device
Cardinal Health 200, LLC
KIT, PRE OP DRAH - Product Usage: Preoperative kit used to decolonize intranasal S. aureus.
Only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment.
Terminated
null
2020-07-30
2020-10-07
null
9,150 kits
null
3010535719
2
0
No Action Indicated (NAI)
2016-08-23
0
0
null
false
0
packaging
incorrect quantity
Z-0284-2021
device
Philips North America, LLC
Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop star...
The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
Terminated
null
2020-09-14
2020-11-04
null
8,942
null
1218950
6
4
Official Action Indicated (OAI)
2018-12-04
2
1
2017-10-31
true
0
device_malfunction
paddle failure detection
Z-1921-2012
device
DePuy Orthopaedics, Inc.
LUSTER HIP SZ 7 Packaging: The product was packaged within inner and outer blister trays, sealed with coated Tyvek lidding. The stems were packaged within vinyl protectors, with foam taper protectors. Intended Use: The LUSTER Cemented Polished Femoral Hip Stems product are forged CoCr polished cemented stems whic...
DePuy Orthopaedics, Inc. is issuing a Field Safety Notice for the LUSTER Cemented Polished Femoral Hip Stems and the ULTIMA LX Cemented Polished Femoral Hip Stems that were implanted with any metal-on-metal (MoM) articulation (including a head of CoCr and an acetabular cup where the cup is of an all metal construction ...
Terminated
null
2012-05-16
2012-07-11
null
(85) LUSTER Cemented Polished Femoral Hip Stems in 69 lots
null
3005572635
1
0
No Action Indicated (NAI)
2009-01-12
0
0
null
false
0
device_malfunction
metal-on-metal articulation
D-1174-2015
drug
Lincare, Inc.
Fluorouracil 4800mg / D5W 296mL, Qty: 1 296 mL Bag, Rx Only. Compounded by: United Medical IV Center (Little Rock) 1527 South Bowman Rd. Ste. D Little Rock, AR 72211-4200.
Lack of Assurance of Sterility: Sterility of product is not assured.
Terminated
null
2015-05-28
2015-07-08
United States including: Arkansas
1 Bag
null
3007636455
1
0
No Action Indicated (NAI)
2012-04-24
0
0
null
false
0
sterility
lack of sterility assurance
Z-1265-2020
device
Abbott Point Of Care Inc.
i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be used for the quantitative measurement of pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the blue i-STAT G3+ cartridges and therefore will no longer distribute the i-STAT G3+ (blue) cartridge in the United States as of May 1, 2020.
Terminated
null
2020-01-14
2020-02-26
null
4,448,525
null
2245578
2
0
No Action Indicated (NAI)
2017-09-20
0
0
null
false
0
device_malfunction
insufficient clinical evidence
Z-0512-2018
device
Philips Medical Systems (Cleveland) Inc
BrightView Model 882480 Product Usage: BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as c...
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
Terminated
null
2017-12-15
2018-02-14
null
1387 in total
null
1525965
10
2
No Action Indicated (NAI)
2019-11-26
1
1
2011-06-07
false
-1
device_malfunction
collimator alignment failure
Z-0099-2026
device
Boston Scientific Corporation
Model Number S702, ALTRUA 2 DR SL Pacemaker
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life ...
Ongoing
null
2025-08-20
2025-10-22
null
4055 units
null
1000120754
3
0
Voluntary Action Indicated (VAI)
2014-10-09
0
0
null
false
0
software
safety mode initiation
D-1201-2015
drug
Lincare, Inc.
Ceftazidime 2 gm / NS 50mL, Qty: 15 50 mL Bags, Rx Only. Compounded by: United Medical IV Center (Little Rock) 1527 South Bowman Rd. Ste. D Little Rock, AR 72211-4200.
Lack of Assurance of Sterility: Sterility of product is not assured.
Terminated
null
2015-05-28
2015-07-08
United States including: Arkansas
15 Bags
null
3002649792
1
0
No Action Indicated (NAI)
2012-03-22
0
0
null
false
0
sterility
sterility not assured
Z-1654-2018
device
Siemens Healthcare Diagnostics, Inc.
Dimension Thyroxine (T4) assay (Product Number: DF65/10444908) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
Terminated
null
2017-06-29
2018-05-16
null
10,056
null
1000660462
1
0
No Action Indicated (NAI)
2013-04-03
0
0
null
false
0
mislabeling
assay interference undisclosed
Z-1828-2015
device
Terumo Cardiovascular Systems Corp
Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.
The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.
Terminated
null
2015-05-06
2015-07-01
null
163 devices
null
1000149028
3
0
Voluntary Action Indicated (VAI)
2017-09-25
0
0
null
false
-1
device_malfunction
incorrect indication use
Z-1495-2024
device
ConvaTec, Inc
ConvaTec CarboFlex, 15cm x 20cm, 6 in. x 8 in., Order No. 403204, Odor Control Dressing, Sterile. Packaged 1 dressing/immediate package, 5 packages/market unit.
The product does not meet sterility assurance level.
Ongoing
null
2024-02-27
2024-04-17
null
63 market units
null
3011987967
6
2
No Action Indicated (NAI)
2015-09-08
1
1
2013-05-24
false
0
sterility
inadequate assurance level
Z-1884-2015
device
Covidien LLC
Palindrome Precision SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 19cm Item Code: 8888119368P
Catheters may not meet antimicrobial specification
Terminated
null
2015-04-23
2015-07-08
null
72,363 US - 24,631 Internationally
null
1221611
1
0
No Action Indicated (NAI)
2010-11-30
0
0
null
false
0
potency
antimicrobial specification failure
D-1256-2022
drug
Nephron Sterile Compounding Center LLC
HYDROmorphone HCl Injection, USP, 6 mg/30 mL (0.2 mg/mL) syringes, 5 x 30 mL Pre-Filled Syringes per carton, 6 x 5 Syringe Carton per case, Rx Only, Nephron 503B outsourcing facility, 4500 12th St. Extension, West Columbia, SC 29172, NDC: 69374-529-30
Lack of Assurance of Sterility
Terminated
null
2022-06-30
2022-07-27
Nationwide in the USA
12,900 syringes
null
null
null
null
null
null
0
0
null
false
null
sterility
lack of assurance
D-0355-2018
drug
Sage Products Inc
Toothette Short Term Oral Swab System with Perox-A-Mint Solution. Contains: 20 Individually wrapped Toothette Plus Oral Swabs with sodium bicarbonate, 1.5 fl.oz/44mL bottle of Perox-A-Mint solution and One (1) 0.5 oz/14g tube of Mouth Moisturizer, Sage Products, Inc., 3909 Three Oaks Road, Cary, IL 60013. Product C...
Cross Contamination With Other Products: Oral care solutions were manufactured by a third party supplier on equipment shared with non-pharmaceutical products
Terminated
null
2017-08-22
2018-02-21
Distributed Nationwide in the USA and Puerto Rico, Canada, United Kingdom, Ireland, Netherlands, New Zealand
966 cases
null
null
null
null
null
null
0
0
null
false
null
contamination
cross contamination
Z-1726-2025
device
Baxter Healthcare Corporation
Welch Allyn Life2000 Compressor, Product Codes: 1) BT-80-0004; 2) BT-80-0004A; 3) BT-80-0008; 4) BT-80-0008A; 5) RMS010093CP.
A cybersecurity vulnerability was discovered through internal testing.
Ongoing
null
2025-04-07
2025-05-21
null
4174 units
null
2025695
3
1
No Action Indicated (NAI)
2017-03-03
1
1
2013-12-19
false
0
software
cybersecurity vulnerability
D-1391-2020
drug
Fresenius Kabi USA, LLC
Fosaprepitant for Injection, 150 mg / vial in a 10 mL Single-Dose vial, Rx only, Fresenius Kabi, Lake Zurich, IL 60047; NDC 63323-972-10
Labeling Error: Label Error on Declared Strength: Carton label and product insert incorrectly states the quantity of the excipient edetate disodium (EDTA) as 5.4 mg / vial, rather than the actual amount of 18.8 mg / vial.
Terminated
null
2020-07-13
2020-07-22
Nationwide within the United States
63,067 vials
63323-972
3014732157
3
2
Official Action Indicated (OAI)
2025-08-08
1
1
2024-01-04
false
-1
mislabeling
excipient quantity error
Z-2495-2018
device
Cardinal Health 200, LLC
Hydroline Trumpet Valve, Trumpet Valve, No Probe, Double Spike, Ref ASU1221, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
Terminated
null
2018-06-11
2018-07-25
null
A total of 63,020 valves were distributed. Specific product quantities were not provided.
null
3034315018
1
0
No Action Indicated (NAI)
2025-09-18
0
0
null
false
-1
device_malfunction
valve closure failure
Z-1288-2022
device
Baxter Healthcare Corporation
#1 MAC F/O LARYNGOSCOPE; F/O Laryngoscope Set-MAC; F/O Laryngoscope Set-MAC W/LED
Packaging error.
Ongoing
null
2022-05-17
2022-07-06
null
99 units
null
2649614
7
1
Voluntary Action Indicated (VAI)
2023-11-29
1
1
2014-07-17
false
-1
packaging
labeling error
Z-2378-2023
device
Covidien LP
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resul...
Ongoing
null
2023-06-28
2023-08-23
null
655 units
null
1282497
8
0
Voluntary Action Indicated (VAI)
2022-11-18
0
0
null
false
0
device_malfunction
catheter lumen occlusion
D-1321-2022
drug
Mckesson Medical-Surgical Inc. Corporate Office
Pataday, Once Daily Relief, Olopatadine hydrochloride ophthalmic solution 0.2%, Antihistamine, Eye Allergy Itch Relief, Once Daily, Sterile, 2.5 mL (0.085 FL OZ) bottle per box, Alcon, NDC 0065-8150-01.
CGMP Deviations: products were stored outside the drug label specifications.
Terminated
null
2022-06-29
2022-08-17
Nationwide in the USA.
1 box
0065-8150
null
null
null
null
null
0
0
null
false
null
cGMP
improper storage conditions
D-1462-2015
drug
The Compounding Pharmacy of America
Dexpanthenol 250 mg/ml Injectable, ml Multi-Dose Vial, Sterile, Refrigerate, The Compounding Pharmacy of America, Knoxville, TN
Lack of Assurance of Sterility; FDA inspection identified GMP violations potentially impacting product quality and sterility
Terminated
null
2015-06-02
2015-09-23
Nationwide
262 mL total volume
null
null
null
null
null
null
0
0
null
false
null
sterility
GMP violations
Z-0090-2026
device
Boston Scientific Corporation
Model Number L221 PROPONENT DR EL Pacemaker
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life ...
Ongoing
null
2025-08-20
2025-10-22
null
3592 units
null
3001451463
6
0
No Action Indicated (NAI)
2016-04-08
0
0
null
false
0
software
safety mode prevention
Z-2487-2018
device
Cardinal Health 200, LLC
Hydroline Trumpet Valve with Pulse Wave Cassette, Trumpet Valve with Cassette, 5mm x 33cm Composite Probe, Double Spike, Ref ASC1201, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
Terminated
null
2018-06-11
2018-07-25
null
A total of 63,020 valves were distributed. Specific product quantities were not provided.
null
2016452
3
0
No Action Indicated (NAI)
2021-12-21
0
0
null
false
-1
device_malfunction
valve closure failure
Z-1392-2019
device
Boston Scientific Corporation
AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vapo...
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
Terminated
null
2019-04-08
2019-05-29
null
27 (OUS)
null
1000120754
3
0
Voluntary Action Indicated (VAI)
2014-10-09
0
0
null
false
0
device_malfunction
cooling system degradation
D-1222-2022
drug
TMC Acquisition LLC dba Tailor Made Compounding
Procaine HCL 2% (20mg/mL), 30 ml-vial, Refrigerate, Tailor Made Compounding
Lack of Assurance of Sterility
Ongoing
null
2022-03-07
2022-07-20
Nationwide USA and Puerto Rico
58 vials
null
null
null
null
null
null
0
0
null
false
null
sterility
sterility assurance failure
Z-2388-2024
device
B. Braun Medical, Inc.
Infusomat UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363421
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
Ongoing
null
2024-06-17
2024-07-24
null
1325700
null
3014301563
1
0
Voluntary Action Indicated (VAI)
2018-12-27
0
0
null
false
0
device_malfunction
backcheck valve failure
D-1695-2019
drug
RXQ Compounding LLC
PREDNISOLONE 1%, MOXIFLOXACIN 0.5% OPHTH SOL, 5ML per dropper bottle, RXQ Compounding, 340 West State Street, Unit 9, Athens, OH 45701. NDC: 70731-0109-05
Lack of Assurance of Sterility
Terminated
null
2019-06-19
2019-08-14
Nationwide USA
N/A
null
3011509553
2
2
Official Action Indicated (OAI)
2017-08-08
1
1
2016-05-19
true
0
sterility
sterility assurance failure
Z-0971-2018
device
Baxter Healthcare Corporation
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) (GTIN 00085412091570); Model Numbers: 35700BAX and 35700ABB. Intended to be used for the controlled administration of fluids.
Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
Terminated
null
2017-10-06
2018-03-21
null
173,375
null
2650090
9
1
Voluntary Action Indicated (VAI)
2025-03-26
1
1
2011-01-20
false
-1
device_malfunction
pump mechanism wear
Z-1277-2025
device
Smiths Medical ASD, Inc.
smiths medical ProPort Plastic Venous Access System, Silicone Catheter, 2.87,,O.D. (8.4Fr) s 1.0mm I.D., REF 21-4150-24
The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect.
Ongoing
null
2025-02-13
2025-03-26
null
3 units
null
1221261
3
0
Voluntary Action Indicated (VAI)
2013-12-10
0
0
null
false
0
device_malfunction
port separation defect
Z-1007-2017
device
Medtronic Navigation, Inc.
AxIEM Touch N Go Pointer Probe. Model Number: 9735318. Placed in a 30mil PETG tray sealed with a Tyvek lid and then placed in a shelf carton.
Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStation¿ System with Touch N Go Pointer Probes for patient registration.
Terminated
null
2016-12-12
2017-01-18
null
111 devices
null
1000517638
11
1
Voluntary Action Indicated (VAI)
2024-02-09
1
1
2010-05-07
false
-1
device_malfunction
registration accuracy issue
Z-0257-2022
device
CooperSurgical, Inc.
LEEP PRECISION Integrated System, 120V CooperSurgical part number: LP-10-120
Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.
Ongoing
null
2021-09-29
2021-11-24
null
1587 Units
null
1216677
15
1
Voluntary Action Indicated (VAI)
2024-05-21
2
2
2021-02-12
false
-1
device_malfunction
coagulation function failure
Z-1263-2023
device
DePuy Orthopaedics, Inc.
Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Ongoing
null
2023-02-15
2023-03-29
null
7 units
null
1219655
2
0
No Action Indicated (NAI)
2015-09-24
0
0
null
false
0
device_malfunction
excessive irradiation dose
Z-0786-2013
device
Philips Medical Systems (Cleveland) Inc
Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Scanner are computed Tomography X-ray systems with multi-slice capability of up to 16 slices,which are intended to produce cross-sectional images of the body by computer reconstruc...
The low contrast pins in the Image Quality (IQ) phantom deteriorates over time. This prevents the customers from doing the Daily Image Quality Checks as defined in the current Instructions for Use (IfU).
Terminated
null
2012-09-18
2013-03-06
null
1,474 Units
null
3018004899
1
0
No Action Indicated (NAI)
2025-01-16
0
0
null
false
-1
device_malfunction
phantom pin degradation
Z-1014-2020
device
Cardinal Health
ROYALSILK 3XLARGE/XLONG NS, Gown Item Code 95998N
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
Terminated
null
2020-01-11
2020-02-12
null
7,659,634 total units
null
1017768
3
0
Voluntary Action Indicated (VAI)
2016-05-12
0
0
null
false
0
contamination
bioburden and particulates
Z-2619-2018
device
ICU Medical, Inc.
147" (373 cm) Bifuse Blood Set w/170 Micron Filter, Hand Pump, 2 Clave¿ w/Remv 3 Gang 1o2¿ Manifold (Blue, Yellow, Red), Check Valve Rotating Luer, 1 Ext, Sterile The blood tubing accessory with CLAVE Connector is a single use, sterile, non-pyrogenic device intended for use as an adjunct to blood tubing sets or veni...
Some devices were labeled as Does not contain DEHP even though the product may contain small amounts of DEHP.
Terminated
null
2017-07-31
2018-08-15
null
160,016 units total
null
3013319212
8
2
No Action Indicated (NAI)
2024-09-09
2
2
2013-05-09
false
-1
mislabeling
DEHP content claim
Z-1822-2017
device
Philips Electronics North America Corporation
The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper F...
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, ...
Terminated
null
2017-03-22
2017-04-26
null
6992 (total distributed for all products)
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
cooling system leak
Z-1140-2019
device
Philips Medical Systems Nederlands
Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive pro...
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
Terminated
null
2019-02-19
2019-04-24
null
7209 total
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
footswitch pedal bending
Z-2189-2021
device
Heartware, Inc.
HeartWare HVAD Pump Implant Kit, REF 1104
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as c...
Ongoing
null
2021-06-03
2021-08-18
null
943 units
null
3010767211
2
1
Voluntary Action Indicated (VAI)
2017-07-21
0
0
null
false
0
device_malfunction
pump failure risk
Z-1013-2019
device
Boston Scientific Corporation
Flexima Duodenal Bend Biliary Stent with Delivery System, 7F (2.3mm) x7cm, Material Number M00539210
The sterile barrier may contain packaging seal defects.
Terminated
null
2018-12-26
2019-03-20
null
18 units
null
1051710
1
0
No Action Indicated (NAI)
2012-02-29
0
0
null
false
0
null
null
D-0532-2023
drug
Pfizer Inc.
MEKTOVI (binimetinib) tablets, 15 mg, 180-count bottle, Rx only, Distributed by: Array BioPharma Inc., a wholly owned subsidiary of Pfizer Inc., Boulder, CO 80301. NDC: 70255-010-02
Labeling: Incorrect or Missing Lot and/or Exp Date: The carton and bottle labels state an expiry date of March 2026; the correct expiration date is February 2025.
Ongoing
null
2023-03-08
2023-04-19
Nationwide in the USA
1,926 Bottles
70255-010
2416101
5
1
No Action Indicated (NAI)
2022-07-22
2
2
2012-05-24
false
0
mislabeling
incorrect expiration date
Z-1925-2015
device
Roche Diagnostics Operations, Inc.
HDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of HDL-cholesterol concentration in serum and plasma.
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administra...
Terminated
null
2015-05-07
2015-07-08
null
6255
null
1823260
9
0
No Action Indicated (NAI)
2022-12-07
0
0
null
false
-1
software
assay interference
Z-0690-2021
device
Boston Scientific Corporation
Five IceRod" i-Thaw" Prostate Cryoablation Kit
Complaint trend regarding needle shaft gas leaks.
Completed
null
2020-11-18
2020-12-30
null
178 units
null
3010178326
1
0
Voluntary Action Indicated (VAI)
2014-03-20
0
0
null
false
0
device_malfunction
needle shaft leaks
Z-0549-2013
device
Zimmer, Inc.
Persona, The Personalized Knee System, Tibial Broach, Cemented, Size G-H, Nonsterile, This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosi...
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone f...
Terminated
null
2012-11-14
2012-12-26
null
92
null
1000220733
10
4
No Action Indicated (NAI)
2022-07-27
0
0
null
false
-1
device_malfunction
tibial perforation risk
Z-2632-2014
device
Fresenius Medical Care Holdings, Inc.
BD Vacutainer Trace Element Tubes (ROYAL BLUE TOP TUBE 6.0ML) Part Number: 363083 Part Number: FMC 87-8380-1 Single use tube used to collect, separate, transport or process blood specimens for laboratory determinations for in vitro diagnostic use. Trace elements, aluminum and zinc determination.
Products were held at improper storage temperature
Terminated
null
2014-06-23
2014-09-24
null
2 each
null
3001451489
8
4
Official Action Indicated (OAI)
2024-10-16
3
3
2023-12-04
false
-1
stability
improper storage temperature
Z-1016-2020
device
Cardinal Health
AAMI 3 NONREIN SCRBNURGOWN BNS L, Item Code 90270NB
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
Terminated
null
2020-01-11
2020-02-12
null
7,659,634 total units
null
3001451686
3
0
No Action Indicated (NAI)
2017-11-21
0
0
null
false
0
contamination
bioburden and particulates
D-1563-2012
drug
Franck's Lab Inc., d.b.a. Franck's Compounding Lab
LIDOCAINE/TETRACAINE OPHTHALMIC 4%/0.5% SOLUTION 10 ML 15 ML 20 ML 200 ML 240 ML 300 ML 400 ML (7 DIFFERENT PRODUCTS)
Lack of Assurance of Sterility: Franck's Lab Inc. initiated a recall of all Sterile Human Drugs distributed between 11/21/2011 and 05/21/2012. FDA environmental sampling revealed the presence of microorganisms and fungal growth in the clean room where sterile products were prepared.
Terminated
null
2012-05-21
2012-08-22
Nationwide, Bahamas, Columbia, Dominican Republic, Grand Cayman, Guatemala, Poland, Santo Domingo, Venezuela, West Indies
8 units
null
null
null
null
null
null
0
0
null
false
null
sterility
microorganism contamination
D-0845-2022
drug
Mylan Pharmaceuticals Inc
Insulin Glargine (insulin glargine-yfgn) Injection, 100 units/mL (U-100), packaged in a 10 mL Multiple-Dose vial inside a carton, Rx only, Manufactured by: Mylan Pharmaceuticals Inc., Morgantown, WV 26505, Manufactured for: Mylan Specialty L.P., Morgantown, WV 26505, NDC 49502-393-80
Labeling: Missing label on the vial
Terminated
null
2022-04-12
2022-05-04
USA nationwide.
89,665 vials
null
1051627
1
0
No Action Indicated (NAI)
2014-04-16
0
0
null
false
0
mislabeling
missing label
Z-2451-2016
device
Biomet 3i, LLC
Titanium Square Uniscrew Dental implants
Pouches may not have been sealed during packing.
Terminated
null
2015-11-25
2016-08-24
null
1,648,273 devices (all products subject to this recall)
null
null
null
null
null
null
0
0
null
false
null
packaging
unsealed pouches
Z-1164-2022
device
Busse Hospital Disposables, Inc.
Pain Tray Catalog Number: 9883
Busse Convenience kits contain one or more swab/swabstick drug products, manufactured by Professional Disposables International (PDI), that are subject to a drug recall due to uncertainty of the adequacy of the validation of the test methods used to manufacture the drug products
Terminated
null
2022-02-22
2022-06-08
null
421 units
null
2433012
6
0
Voluntary Action Indicated (VAI)
2018-04-13
0
0
null
false
0
cGMP
validation inadequacy
Z-1859-2013
device
Siemens Medical Solutions USA, Inc
syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo¿ Dynamics system is intended for the acceptance, transfer, display, storage, archive, and manipulation of digital medical images, including quantification and reportgeneration. syngo¿ Dynamics is not intended to be used for reading mammogr...
Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the mod...
Terminated
null
2013-06-10
2013-08-07
null
9
null
2240869
3
0
Voluntary Action Indicated (VAI)
2022-10-26
0
0
null
false
-1
software
data save failure
Z-2696-2020
device
Philips North America, LLC
IntelliVue MX800 patient monitor, model no. 865240
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Terminated
null
2020-06-30
2020-07-29
null
28
null
1218950
6
4
Official Action Indicated (OAI)
2018-12-04
2
1
2017-10-31
true
0
device_malfunction
defective power supply
Z-0530-2025
device
Boston Scientific Corporation
AMS Artificial Urinary Sphincter Control Pump
Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
Ongoing
null
2024-10-28
2024-12-04
null
16 units
null
3003882783
1
0
No Action Indicated (NAI)
2018-08-30
0
0
null
false
0
mislabeling
wrong device packaged
Z-0318-2014
device
Ortho-Clinical Diagnostics
VITROS Chemistry Products Mg Slides packaged as 300 slides/pack catalog number 8255093 and packaged as 90 slides/pack catalog number 1921204 VITROS Mg Slides quantitatively measure magnesium (Mg) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 In...
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Terminated
null
2013-06-12
2013-11-27
null
101,976 units (39,909 units domestically & 62,067 units internationally)
null
3007111389
5
0
No Action Indicated (NAI)
2018-02-01
0
0
null
false
-1
device_malfunction
slide dispense failure
D-0407-2021
drug
Cardinal Health Inc.
FENOFIBRATE CAPSULES, UPS 67MG 100 CAPSULES NDC/UPC 27241-118-04; RX; CAPSULES
CGMP Deviations: Intermittent exposure to temperature excursion during storage.
Terminated
null
2021-03-15
2021-06-02
FL, GA, SC
106 BOXES
27241-118
2000029340
2
0
No Action Indicated (NAI)
2021-11-03
0
0
null
false
-1
cGMP
temperature excursion
Z-1136-2020
device
Abbott Vascular
NC TREK Coronary Dilatation Catheter, REF 1012455-08 Rx Only, NC TREK RX 5.00 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation o...
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
Terminated
null
2020-01-29
2020-02-26
null
Total 39687 NC TREK units (13891 US and 25796 OUS)
null
3006169760
1
0
No Action Indicated (NAI)
2011-07-20
0
0
null
false
0
device_malfunction
balloon deflation failure
Z-1517-2024
device
Boston Scientific Corporation
EndoVive 20 Fr Safety PEG Push w/ENFit Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00509191
Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, lea...
Ongoing
null
2024-02-20
2024-04-17
null
13 units (Japan)
null
2648729
5
0
No Action Indicated (NAI)
2021-07-21
0
0
null
false
0
device_malfunction
blocked lumen connector
Z-0876-2019
device
Philips Medical Systems Nederlands
Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Terminated
null
2018-12-21
2019-03-06
null
11226 total
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
magnetic attraction hazard
Z-1962-2014
device
Siemens Medical Solutions USA, Inc.
ACUSON S Family Ultrasound Systems at software versions VC25D, VC30A and VC30B; (S2000, S3000 and S1000 systems) Siemens Medical Solutions USA, Inc. Ultrasound imaging systems.
There is a potential measurement error with the Auto-Stats (auto statistics) measurement tool on the ACUSON S Family ultrasound system.
Terminated
null
2014-06-03
2014-07-09
null
790 units
null
1000514697
3
0
No Action Indicated (NAI)
2019-06-10
0
0
null
false
-1
software
measurement tool error
Z-1834-2021
device
DeRoyal Industries Inc
DeRoyal Angio Cath Removal Tray, REF 89-9936.02
DeRoyal manufactured certain lots of surgical procedure pack products using Hospira Pfizer 1% Lidocaine which was subsequently recalled due to mislabeling; the product is 0.5% Bupivacaine
Terminated
null
2021-05-14
2021-06-16
null
60 trays
null
1043214
4
0
No Action Indicated (NAI)
2017-12-08
0
0
null
false
0
mislabeling
wrong drug concentration
Z-1804-2023
device
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Ongoing
null
2023-05-10
2023-06-28
null
34350 units
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
reduced energy output
Z-1280-2018
device
Baxter Healthcare Corporation
Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. *Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more w...
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentio...
Terminated
null
2017-02-15
2018-04-11
null
3,440 units
null
1025114
11
5
Official Action Indicated (OAI)
2024-02-09
0
0
null
false
-1
software
syringe pump parameter revert
Z-1810-2018
device
Tosoh Bioscience Inc
Tosoh ST AIA-PACK HbA1c Calibrator, HbA1c Calibrator, PN 025318 The product is intended for In Vitro Diagnostic Use Only for the calibration of the HbA1c Assay.
The assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.
Terminated
null
2018-03-05
2018-05-23
null
1485
null
3005529799
6
2
Voluntary Action Indicated (VAI)
2019-10-11
2
2
2016-08-05
false
-1
software
inaccurate results
Z-1773-2015
device
Philips Medical Systems (Cleveland) Inc
Ingenuity CT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Terminated
null
2015-03-06
2015-06-17
null
156 units
null
1525965
10
2
No Action Indicated (NAI)
2019-11-26
1
1
2011-06-07
false
-1
device_malfunction
battery failure hazard
Z-0847-2015
device
Natus Medical Incorporated
Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.
Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some...
Terminated
null
2014-11-19
2014-12-31
null
1907 total (1413 in US and 494 international)
null
2023988
3
0
No Action Indicated (NAI)
2016-04-29
0
0
null
false
-1
device_malfunction
excessive light intensity
Z-0822-2025
device
Boston Scientific Corporation
ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Ongoing
null
2024-12-12
2025-01-22
null
156,329 units (59,937 US, 96,792 OUS) in total
null
3002095335
11
0
No Action Indicated (NAI)
2025-03-18
0
0
null
false
-1
device_malfunction
battery impedance defect
Z-0229-2014
device
Carestream Health Inc.
Carestream Vue PACS; UNLIMITED READING LIC ADD 25K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all ...
Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.
Terminated
null
2013-09-16
2013-11-20
null
null
null
null
null
null
null
0
0
null
false
null
software
image loading failure